Product information "EC Epsilon EClipse™ Potentiostat & Bipotentiostat | EF-1031F"

BASi – EC Epsilon EClipse™ – Potentiostat/Galvanostat/Bipotentiostat
Potential range ±12 V, current range ±100 mA

High-performance electrochemical measurement system for precise process control

The EC Epsilon EClipse™ of BASi Research is a high-performance electrochemical measurement system for users who require precise control and evaluation of electrochemical processes. It combines potentiostat and galvanostat functionality with true bipotentiostat operation, providing high flexibility for demanding laboratory and development tasks.

Electrochemical measurement system for research and development

The Epsilon EClipse™ supports a wide range of established electrochemical methods and is suitable for both routine applications and complex experimental setups. Hardware and software are designed to deliver reproducible measurement results and controlled execution of demanding experiments.

Wide range of methods

Supported techniques include cyclic and linear sweep voltammetry, chronoamperometry, and chronopotentiometry. Extended software packages additionally enable methods such as differential pulse, square wave, and stripping voltammetry. Programmable sequences automate measurement procedures and improve repeatability.

True bipotentiostat operation

Two independently controlled working electrodes enable bipotentiostatic measurements and parallel experiments on separate channels. This allows targeted investigation of complex electrochemical systems and multi-electrode configurations.

Software and integration

The PC software supports structured method control and clear visualization of measurement data. Users define triggers, delays, and repetitions and integrate the Epsilon EClipse™ efficiently into existing laboratory workflows.

Typical applications

The EC Epsilon EClipse™ is used in electrochemical research, analytical chemistry, corrosion studies, battery research, and electrode and coating investigations. Wherever current and potential must be precisely controlled and measured, the system provides a reliable solution.

Request consultation

Would you like to use the EC Epsilon EClipse™ in your electrochemical measurement application? Contact us – we will advise you on the appropriate configuration and support you in selecting the optimal solution.

Current: < 1 A
Functionality: Bipotentiostat, Galvanostat, Impedance Analyzer, Potentiostat
Output Channels: 1

Related links to "EC Epsilon EClipse™ Potentiostat & Bipotentiostat | EF-1031F"

Available documents (Subject to change)

Specifications / Features of the Epsilon EClipse™ EF-1031F

  • Electrochemical operating modes:
    • Potentiostat
    • Galvanostat
    • Potentiometer
    • Bipotentiostat option
    • Wide current range from microelectrodes to small-scale electrosynthesis
  • Operation & software:
    • Multiple Document Interface (MDI)
    • Multiple experimental runs
    • File overlay
    • File subtraction
    • Multiple text data file formats, including DigiSim
    • Automatic and manual peak detection
    • Initiation at open-circuit potential
    • IUPAC or polarographic axis convention
  • Signal conditioning & noise reduction:
    • Optically isolated circuitry for noise reduction
    • Analog filtering
    • Digital smoothing (moving average)
    • 50/60 Hz selection for mains noise minimization
    • Positive-feedback iR compensation with stability testing
  • Analysis functions:
    • Anson plot (Q vs. t1/2) for chronocoulometry
    • Cottrell plot (i vs. t-1/2) for chronoamperometry
  • System test & automation:
    • Internal dummy cells for hardware testing
    • Remote control of BASi® cell stands (CGME, C-3 Cell Stand, RDE-2)

Potentiostat Specifications

  • Channel 1 – applied potential
    • Dynamic DAC: 16 bit
    • ±3.275 V at 0.1 mV resolution
    • ±10 V at 0.33 mV resolution
    • Offset DAC: 8 bit, ±2.55 V at 10 mV resolution
  • Channel 2 – applied potential
    • Dynamic DAC: 8 bit
    • ±2.55 V at 10 mV resolution
  • Electrical characteristics (both channels)
    • Compliance voltage: ±12 V
    • Maximum current: 100 mA
    • Bandwidth: > 100 kHz
    • Bias current: < 1 pA
    • Electrometer input impedance: > 1012 Ω

Options and Accessories of the Epsilon EClipse™ EF-1031F

  • EF-1085: BASi C-3 Cell Stand
  • EF-1100: BASi RDE-2 Cell Stand
  • EF-1400: BASi CGME Cell Stand
  • MF-8285: Laptop with preinstalled software

Scope of Delivery of the Epsilon EClipse™ EF-1031F

  • Software:
    • Basic Software
    • Basic Plus Software
    • Methods Software
    • Bipotentiostat
  • 10" USB-Cable
  • Cell connection cable for bipotentiostat

Possible Applications of the Epsilon EClipse™ EF-1031F

  • Analytical chemistry
  • Electrochemistry
  • Battery test
  • Corrosion test
  • Synthesis
  • Research and development
  • etc.

Founded in 1974, BASi Research Products, Inc is a part of Inotiv, Inc- a publicly traded, global company headquartered in Indianapolis, Indiana. From the beginning, BASi has provided niche instrumentation to the life sciences  and electrochemistry markets. They serve researchers throughout the world with instruments and supplies that help them explore new frontiers in pharmaceutical and academic endeavors.