- Order number: N4-PPA1530
Newtons4th (N4L) PPA1530
Compact Power Analyzer
3 Phases, 20 A Shunt, DC~1 MHz Bandwith
The PPA1500 Power Analyzer offers higher performance than its sibling – the PPA500 unit, the PPA1500 series ensures that N4L power analyzers provide the optimum solution in almost any power measurement application and budget. The main differences between the PPA500 and PPA1500 Power Analyzers are that the PPA1500 features a 1Mhz Bandwidth and full graphical display, this includes harmonic bar graphs and graphical data logging. The PPA1500 Power Analyzers now features sophisticated parallel digital filtering algorithms enabling the PPA1500 to be used in various Pulse Width Modulated applications. While being only half rack width and 2U high, the PPA1500 is available in 1, 2 or 3 phase configurations and offers a level of performance normally found only in larger and more costly instruments. With modes such as “Inrush Current and “Standby Power”, the PPA1500 provides fast, accurate and easy measurements in challenging environments.
The PPA1500 series analyzers now have a 6 Phase 2U Rack option in which two PPA1530 units are installed on the same rack side by side
All PPA1500 units are supplied with UKAS ISO17025 certification as standard, this is performed at N4L in the UK in our accredited ISO17025.
- max Input voltage: 2,5 kVpk (1kVrms)
- max Input current: 20 A or 30 A
- Phases / Channel: 1-3
- Bandwidth: DC~1 MHz
- Accuracy V, A rdg Error: 0,05 %
- Accuracy Power rdg Error: 0,10 %
Features of Compact Power Analyzer PPA1530
- max Input voltage: 2,5 kVpk (1kVrms)
- max Input current: 20 A
- Phases / Channel: 3
- Bandwidth: DC~1 MHz
- Accuracy V, A rdg Error: 0,05 %
- Accuracy Power rdg Error: 0,10 %
- Scope and Graph Modes
- PWM FilterOptions: 2
- Speed / Harmonics / Sec Computation: 300/sec
- Internal Dataloging: 4 Paramter
- Datalog Records: 16.000
- Internal Memory: 192 kB
- Harmonics: 50 Minimum
- Minimum Window Size: 5ms
- 0.05% Basic accuracy (Traceable to ISO17025)
- High precision internal shunts Superior accuracy to any external shunt
- 20 Arms (300 Apk) & 1000 Vrms (2500 Vpk) direct input
- High speed sampling on all channels
- True real time analysis with no measurement gap
- Simple BNC connection of N4L shunts for high current applications
- Frequency range DC and 10 mHz to 1 MHz
- 10 milli degree phase accuracy – Necessary for low power factor applications
- Unique x10 gain mode for Standby Power measurement
- High contrast full colour graphic display
- LAN, RS232, USB, Extension and Auxiliary ports / Optional GPIB
- USB Memory Port
Applications of Compact Power Analyzer PPA1530
- PWM Motor Drive Efficiency Measurements
- Standby Power Measurement of a USB Mains Adaptor
- AC/DC Converter Efficiency Measurements
- AC-DC power efficiency measurements of power supplies
Newtons4th Ltd (abbreviated to N4L) was established in 1997 to design, manufacture and support innovative electronic equipment to a world-wide market.
Specialising in sophisticated test equipment for the power electronics industry including Power Analyzers, Frequency Response Analyzers (gain/phase analyzers), AC Power Sources, Impedance Analyzers and Selective Level Meters.