ITECH IT-N2100 | Solar Array Simulator
Evaluate MPPT behavior, efficiency, and function independently of weather and physical PV modules under repeatable conditions.
- ✓ Power: up to 1.5 kW
- ✓ MPPT Tracking: up to 250 Hz
- ✓ I-V Curve: up to 4.096 Data Points
Solar Array Simulator for Microinverters, Power Optimizers, and MPPT Testing
PV Simulator for Static and Dynamic I-V Curves
The ITECH IT-N2100 Series reproduces defined I-V curves for testing microinverters, power optimizers, and the PV inputs of energy storage systems. Irradiance, temperature, and shading can be set as required. This enables comparable MPPT, efficiency, and functional tests under controlled conditions.
Dynamic MPPT Testing
The current control loop maintains the specified I-V curve even at DUT MPPT tracking frequencies of up to 250 Hz. Up to 4,096 data points define static and dynamic curves in high detail and reproduce changes in irradiance, temperature, or shading with precision. Minimal output current ripple keeps the source’s influence on current sampling and MPPT evaluation very low.
Test Models and Multi-Channel Operation
Five PV operating modes cover user-defined curves, list sequences, integrated PV curve models, fixed mode, and table-based curves. Test models in accordance with EN 50530, Sandia, NB/T 32004, CGC/GF004, and CGC/GF035 can be executed and documented via software. USB and LAN enable remote control using SCPI, while up to 20 channels can be operated synchronously.
Selection by PV Input Requirements and Test Power
The IT-N2100 Series includes models for different voltage, current, and power requirements at the DUT’s PV input. Model selection depends on the required operating range and maximum test power. The isolation rating of up to 1,500 V DC extends the use of the IT-N2100 Series to high-voltage test setups.
Alternative Products
For PV simulation with significantly higher voltage, current, or power requirements, consider the Kewell D2000-IV PV Simulator. For test systems that also require bidirectional source and sink operation with regenerative feedback, the ITECH IT6000B Series with solar array simulation software is a suitable alternative.
Related Product Categories
Simulators | Power Supplies and Sources | Bidirectional Power Supplies
Configure PV Testing Around Your DUT
Check whether the voltage range, power, dynamics, and channel count match your DUT’s PV input. We support you in selecting, configuring, and integrating the IT-N2100 Series.
Variants
Specifications
| Product class | solar array simulator and programmable DC power supply for simulating photovoltaic I-V curves |
|---|---|
| Models | IT-N2121, IT-N2123, IT-N2131, and IT-N2133 |
| Output power | model-dependent, 800 W or 1,500 W |
| Output voltage | model-dependent, 0 V to 80 V or 0 V to 150 V |
| Output current | model-dependent, up to 10 A or 25 A |
| 80 V models | output voltage expandable up to 85 V |
| Wide-range output | available on the 800 W models for different voltage and current combinations within the power limit |
| Current control loop | bandwidth up to 500 kHz |
| Supported MPPT tracking frequency | up to 250 Hz |
| Dynamic voltage response | model-dependent, 150 µs to 400 µs |
| Programming response time | ≤ 15 ms |
| I-V curves | up to 4,096 data points |
| PV operating modes | user-defined mode, curve mode, List Mode, fixed mode, and table mode |
| Four-point method | rapid curve generation using Voc, Isc, Vmp, and Imp |
| Fixed mode | CC, CV, and CP output behavior |
| Output impedance | adjustable from 0 Ω to 9.999 Ω on the 80 V models |
| PV cell models | simulation of I-V curves for silicon, GaAs, and other solar cell technologies |
| PV test models | EN 50530, Sandia, NB/T 32004, CGC/GF 004, and CGC/GF 035 |
| Dynamic PV simulation | up to 65,000 time points with 1 s time resolution |
| List Mode | up to 100 steps in 20 data sets |
| Shading simulation | static and dynamic I-V curves, including moving shading profiles |
| Environmental simulation | I-V curves under varying irradiance and temperature |
| Multi-channel control | synchronous control of up to 20 solar array simulators via the PV software |
| File management | create, upload, download, and store up to 100 I-V and PV files |
| Measurement logging | real-time acquisition of voltage, current, and power |
| Data export | CSV and TDMS files to USB storage devices |
| Report function | automatic generation of standardized test reports |
| Display | 4.3-inch HD LCD for graphical display of curves and MPPT status |
| Voltage resolution | 10 mV for setting and 1 mV for readback |
| Current resolution | 1 mA for setting and readback |
| Voltage setting accuracy | model-dependent, ≤ 0.03% + 20 mV or ≤ 0.03% + 30 mV |
| Current setting accuracy | model-dependent, ≤ 0.05% + 5 mA or ≤ 0.05% + 10 mA |
| Voltage ripple | ≤ 40 mV RMS |
| Current ripple | model-dependent, ≤ 5 mA RMS to ≤ 10 mA RMS |
| Rise and fall time | ≤ 20 ms each at no load and full load |
| Remote sense compensation | ≤ 2 V |
| Common-mode withstand voltage, output to ground | 1,500 V DC |
| Withstand voltage, input to ground | 1,500 V AC |
| AC input | 110 V or 220 V, 50/60 Hz |
| Maximum input current | 11 A |
| Power factor | ≥ 0.98 |
| Typical efficiency | 60% for 80 V models and 72% for 150 V models |
| Protection functions | OVP, OCP, OPP, and OTP |
| Standard interfaces | LAN, USB, and digital I/O; remote control via SCPI |
| Outputs | front- and rear-panel terminals |
| Form factor | 2U half-rack per channel |
| Rack-mount dimensions | 450 × 214 × 88.2 mm |
| Overall dimensions | 529.5 × 255 × 108.2 mm |
| Net weight | 9 kg |
| Cooling | forced-air cooling |
| Operating temperature | 0°C to 40°C |
| Degree of protection | IP20 |
| Safety requirements | IEC 61010 |
Options & Accessories
| IT system cabinet | system cabinet in 15U, 27U, or 37U |
|---|---|
| IT-E30305-IY | test lead pair, 1 m, I-type terminal to fork terminal |
| IT-E30615-OO | high-current cable pair, 60 A, 1.5 m, round terminal to round terminal |
| IT-E30312-YY | test lead pair, 30 A, 1.2 m, fork terminal to fork terminal |
| IT-E30320-YY | test lead pair, 30 A, 2 m, fork terminal to fork terminal |
| IT-E30110-AB | test lead pair, 10 A, 1 m, alligator clip to banana plug |
| IT-E30110-BB | test lead pair, 10 A, 1 m, banana plug to banana plug |
| IT-E30110-BY | test lead pair, 10 A, 1 m, banana plug to fork terminal |
| AC power cord | configuration-dependent mains connection cable |
| IT-E9003 | ITECH warranty extension for up to 3 years |
| IT-E9005 | ITECH warranty extension for up to 5 years |
Scope of Delivery
| IT-N2100 Series unit(s) |
| AC power cord |
| US connection cable |
| Calibration certificate |
Applications
| Static and dynamic MPPT testing of microinverters |
| Testing of PV power optimizers |
| Testing of multi-channel inverters and PV inputs |
| Simulation of solar modules and PV arrays under varying irradiance and temperature |
| Testing inverter behavior under partial shading and dynamic I-V curves |
| Development, verification, and production testing of PV inverters |
| Burn-in and endurance testing of photovoltaic power-conversion systems |
| Testing of PV-coupled home energy storage and power-supply systems |
| Testing of photovoltaic power-supply systems for aerospace applications |